RayonixHs Tango device

This is the reference documentation of the RayonixHs Tango device.

you can also find some useful information about the camera models/prerequisite/installation/configuration/compilation in the RayonixHs camera plugin section.

Properties

Property name Mandatory Default value Description
frame_mode No single The frame mode, single or fast_transfer
frame_trigger_signal_type No opto The frame trigger signal type (input #1)
sequence_gate_signal_type No opto The gate signal type (input #2)
electronic_shutter_enabled No false The electronic shutter true or false to activate or not
cooler_temperature_setpoint No -120 The cooling system temperature setpoint in Celsuis
sensor_temperature_setpoint No -80 The detector (sensor) temperature setpoint in Celsuis
output1_signal_type No cmos The output #1 signal type
output2_signal_type No cmos The output #2 signal type
output1_id No shutter The output #1 signal source
output2_id No frame the output #2 signal source
The Rayonix HS input/output system supports different type of signals:
  • OPTO/OPTO_INVERTED/CMOS/CMOS_PULLDOWN/CMOS_PULLUP/CMOS_PULLDOWN_INVERTED/CMOS_PULLUP_INVERTED
And it provides a output multiplexer for both outputs within the following list of sources:
  • SHUTTER/INTEGRATE/FRAME/LINE/SHUTTER_OPENING/SHUTTER_CLOSING/SHUTTER_ACTIVE/TRIGGER_RISE_WAIT/TRIGGER_RISE_ACK/TRIGGER_FALL_WAIT/TRIGGER_FALL_ACK/TRIGGER_2_RISE_WAIT/TRIGGER_2_RISE_ACK/INPUT_FRAME/INPUT_GATE

Attributes

Attribute name RW Type Description
frame_mode rw DevString The frame mode, single or fast_transfer
frame_trigger_signal_type rw DevString The frame trigger signal type (input #1)
sequence_gate_signal_type rw DevString The gate signal type (input #2)
electronic_shutter_enabled rw DevString The electronic shutter true or false to activate or not
cooler_temperature_setpoint rw DevDouble The cooling system temperature setpoint in Celsuis
sensor_temperature_setpoint rw DevDouble The detector (sensor) temperature setpoint in Celsuis
output1_signal_type rw DevString The output #1 signal type
output2_signal_type rw DevString The output #2 signal type
output1_id rw DevString The output #1 signal source
output2_id rw DevString The output #2 signal source
vacuum_valve rw DevString The vacuum valve command true or false to open or close

Warning: be careful with the temperature setting (and vacuum valve), the operating temperature is factory-determined and should never be changed. There is no reason to run the detector at a warner temperature.

For the signal type and source the possible values are listed above in the Properties section.

Commands

Command name Arg. in Arg. out Description
Init DevVoid DevVoid Do not use
State DevVoid DevLong Return the device state
Status DevVoid DevString Return the device state as a string
getAttrStringValueList DevString: Attribute name DevVarStringArray: String value list Return the authorized string value list for a given attribute name

Table Of Contents

Previous topic

Prosilica Tango device

Next topic

RoperScientific Tango device

This Page